Measuring Instruments and Errors

Errors and accuracy, PMMC, moving iron, electrodynamometer and rectifier instruments, extension of ranges with shunts and multipliers, loading effect, and instrument transformers.

Measurements are only as good as the instruments and how they are used. GATE asks which instrument suits AC or DC, how to extend ranges, and how errors combine.

Errors

  • Accuracy: closeness to the true value. Precision: closeness of repeated readings to each other. Resolution: smallest detectable change.
  • Types: gross (human), systematic (instrumental, environmental, loading) and random errors.
  • Limiting (guarantee) error is quoted as a percentage of full-scale, so the relative error grows at lower readings: a ±1% of full-scale instrument reading at half-scale has ±2% error.
  • Combining errors: for a sum or difference, absolute errors add; for a product or quotient, relative (percentage) errors add; for a power xⁿ, n times the relative error.

Analog instruments

InstrumentWorks onScaleReads
PMMCForce on a coil in a permanent magnet fieldLinearDC only (average value)
Moving iron (attraction, repulsion)Iron vane in a coil's field, T ∝ I²Non-linear (cramped at the start)AC and DC (RMS)
ElectrodynamometerFixed and moving coilsNon-linear for V and I; linear as a wattmeterAC and DC, transfer standard
Rectifier typePMMC + diode bridgeCalibrated in RMS for sine wavesAC (errors for non-sinusoidal waves)
Thermocouple / hot-wireHeating effectSquare lawTrue RMS, high frequency

Damping: eddy-current damping (PMMC), air-friction damping (moving iron). Controlling torque: springs or gravity.

Extending ranges

  • Ammeter shunt: R_sh = R_m I_m/(I − I_m) = R_m/(m − 1), with m = I/I_m.
  • Voltmeter multiplier: R_se = V/I_m − R_m = R_m(m − 1).
  • Voltmeter sensitivity (Ω/V) = 1/I_fs. A low-sensitivity voltmeter loads a high-resistance circuit and reads low.

Instrument transformers

  • CT (current transformer): steps current down (for example 1000/5 A). Its secondary must never be opened while the primary carries current: dangerously high voltage and core saturation.
  • PT (potential transformer): steps voltage down (for example 11 kV/110 V).
  • Errors: ratio error and phase-angle error.

Worked example

A PMMC meter with 1 mA full-scale and 100 Ω resistance is to read 10 V.

  • Multiplier = 10/0.001 − 100 = 9900 Ω. Sensitivity = 1000 Ω/V.

Practice

GATE-styleNAT · numerical answer

A 0–100 V voltmeter has a guaranteed accuracy of 2% of full scale. What is the possible percentage error when it reads 50 V?

Show answer and explanationAnswer: 4

Error = ±2 V; at 50 V this is ±4%.

GATE-styleNAT · numerical answer

A meter with R_m = 50 Ω and full-scale current 1 mA must measure 1 A. What shunt resistance (mΩ) is needed? (Give 2 decimal places.)

Show answer and explanationAnswer: 50.05

R_sh = R_m I_m/(I − I_m) = 0.05/0.999 = 0.05005 Ω = 50.05 mΩ.

GATE-styleNAT · numerical answer

Power is found as P = V I. V has ±1% error and I has ±2% error. What is the maximum percentage error in P?

Show answer and explanationAnswer: 3

For a product, relative errors add: 1% + 2% = 3%.

GATE-styleMCQ · one correct answer

Which instrument can measure only DC?

Show answer and explanationAnswer: A

The PMMC torque reverses with current direction, so it reads the average, which is zero for AC.